User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
129
Prepare syringes for emptying the Pump
Syringe 1 is used for emptying the Pump. Prepare syringe 1 as described in the
"Precautions and General Procedures" section in Appendix D of this manual.
Label the syringe per table 2.
After the syringe is prepared, weigh the syringe and enter the weight on the Refill
Form.
Prepare syringes for filling the Pump
During the Catheter flush procedure, syringes 2, 3, 5, and 6 are used for filling
the Pump. Each syringe should be labeled per Table 2, then filled and degassed
following the General Procedures section in Appendix D of this manual.
Prepare syringe for flushing the Side Port Catheter
During the Catheter flush procedure, syringe 4 is used for flushing the Side Port
Catheter. This syringe needs to be labeled and then prepared by following the
steps below:
1. Firmly attach a stopcock to the 10 or 20 ml Luer Lock connector of a 10 or 20
ml syringe. Then attach an 18 gauge needle to the stopcock.
NOTE: Use only a 10-20 ml Luer Lock syringe for this
procedure. Smaller syringes may damage the Catheter
and/or Pump, by allowing too much high pressure.
2. Draw 5-10 mL of RINSE BUFFER into the syringe.
3. Expel all air from the syringe.
4. Close the stopcock and remove the 18 gauge needle from the syringe.
5. Fill the hub of the Medtronic MiniMed refill needle with RINSE BUFFER
from the syringe and attach it to the syringe.
6. Prime the Medtronic MiniMed refill needle completely.
7. Close the stopcock.
8. Set the syringe aside.