User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
127
APPENDIX F Side Port Catheter flush
procedure
When delivery of insulin is impaired due to catheter tip obstruction, the Catheter
may be flushed using 5 - 10 ml of RINSE BUFFER, MMT-2008.
Supplies and solutions
Prior to performing this procedure, assemble the necessary supplies and solutions
as outlined below:
Supplies
• Steri-Strips
®
and markers
• Local anesthesia
• Sharps container
•PPC
• Oral or IV glucose
• Glucose monitoring equipment
•Drapes
• One (1) scale - 0.01 gram resolution
• One (1) sterile bag for PPC (intestinal or cassette bags work well)
• One (1) Side port locating template, MMT-4106
• Five (5) Refill kits, MMT-4105
• Six (6) Medtronic MiniMed MMT-4102 refill needles. (Extra needles
should be available to use as needed)
• Eight (8) 18 gauge regular bevel needles
• One (1) 10 or 20 ml Luer Lock
®
syringe
• One (1) stopcock - 2 way