User's Manual

Table Of Contents
115
APPENDIX E Pump rinse procedure
The purpose of this procedure is to dissolve insulin deposits within the Pump res-
ervoir, pumping mechanism, and the Side Port Catheter “port.”
Supplies and solutions
Prior to performing this procedure, assemble the necessary supplies and solutions
as outlined below:
Supplies
Steri-Strips
®
and markers
Local anesthesia (if necessary)
Sharps container
Safety glasses
•PPC
Oral or IV glucose
Glucose monitoring equipment
•Drapes
One (1) scale - 0.01 gram resolution
One (1) sterile bag for PPC (intestinal or cassette bags work well)
One (1) Side port locating template, MMT-4106
Seven (7) Refill kits, MMT-4105
Seven (7) Medtronic MiniMed MMT-4102 refill needles. (Extra needles
should be available to use as needed)
Eight (8) sharp 18 gauge regular bevel needles