User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
111
Locating the pump fill port and the side port
Figure 15: Template and placement
1. Locate the Pump, then aseptically prep and drape the Pump area. Use topical
or local anesthetic if desired.
2. To locate the Pump fill port, located in the center of the Pump, align the
sterile template (MMT-4106) over the Pump by palpating the Pump
circumference. Then insert an 18 gauge needle into the central Pump fill port
as shown in Figure 15 above.
3. To locate the side port, palpate around the circumference of the Pump. It is
helpful to grasp the side port between the thumb and index finger when
aligning the template.
4. Using the template as a guide, insert the second 18 gauge needle into the side
port as shown. The side port can be accessed by inserting the guide needle
just inside the outer edge of the side port connection. This outer edge is
raised around the outer aspect of the side port, in order to help guide the
needle into the port.
Template
(MMT-4106)
Placement on Skin with
Guide Needles in Position