User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
99
CHAPTER 11 Technical specifications
Implantable Insulin Pump MMT-2007D
Diameter
Thickness
Reservoir Volume
8.1 cm (3.2 inches)
2.0 cm (0.8 inches)
13 ml to 15 ml
Weight - Empty 131 gm (4.6 ounces)
Insulin - Concentration U-400
Stroke Volume 0.42 to 0.58 ul per stroke
0.17 to 0.23 units per stroke
Basal Rate 0.2 to 35.0 units per hour (U/h)
Basal Patterns 3 patterns of up to 48 basal rates each.
Meal Bolus 0.2 to 25.0 units
Bolus Duration Immediate, Square Wave (30 minutes to 4 hours), or Dual Wave.
Audio Bolus
Temporary Basal Rate 0.2 to 35.0 units per hour
30 minute increment duration
30 minutes up to 24 hours delay
Diagnostic Delivery Rate 10 to 150 U/h
Power Supply Lithium - Carbon Monofluoride Battery
Audio Alarms Low Battery
System Error
Safety Features Negative Pressure Reservoir with Passive Filling
Pump Shutdown and Alarm with System Error
(unique code sequences)
Materials in contact with tissue Titanium, Silicone Rubber