Pump suspended
If the Pump operation has been suspended, the following message is displayed:
During “SUSPEND PUMP”, the Pump will deliver a basal rate of
0.2 U/h. To restart insulin delivery programming, press SEL and ACT, then place
the PPC near the Pump and allow the communication to complete.
Auto off
If the “AUTO OFF” time interval elapses in the Pump and the PPC recognizes
this condition, the following message is displayed:
Press SEL and ACT, and then place the PPC near the Pump and allow the com-
munication to complete. The PPC then communicates with the Pump to reset the
“AUTO OFF” duration.
If five more minutes elapse, the Pump will initiate the internal alarm sequence of
a beep every 15 seconds for 10 minutes, then double-beeps every 15 seconds for
10 minutes, then repeating the pattern. The alarm is cleared by pressing SEL and
ACT.
Hourly maximum exceeded
Attempting to deliver more than 2.5 times the pre-programmed bolus maximum
in one hour causes the following message to be displayed:
Press SEL and ACT, and then place the PPC near the Pump and allow the com-
munication to complete. The patient may exceed the pre-programmed bolus limit
by programming another bolus within 10 minutes.
PPC Display Type of Alarm
PUMP
SUSPENDED
2
PPC Display Type of Alarm
AUTO OFF PUMP
PUMP SUSPENDED
1
PPC Display Type of Alarm
HOURLY MAX
EXCEEDED
1