User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
System alarms and messages
95
Pump suspended
If the Pump operation has been suspended, the following message is displayed:
During “SUSPEND PUMP”, the Pump will deliver a basal rate of
0.2 U/h. To restart insulin delivery programming, press SEL and ACT, then place
the PPC near the Pump and allow the communication to complete.
Auto off
If the “AUTO OFF” time interval elapses in the Pump and the PPC recognizes
this condition, the following message is displayed:
Press SEL and ACT, and then place the PPC near the Pump and allow the com-
munication to complete. The PPC then communicates with the Pump to reset the
“AUTO OFF” duration.
If five more minutes elapse, the Pump will initiate the internal alarm sequence of
a beep every 15 seconds for 10 minutes, then double-beeps every 15 seconds for
10 minutes, then repeating the pattern. The alarm is cleared by pressing SEL and
ACT.
Hourly maximum exceeded
Attempting to deliver more than 2.5 times the pre-programmed bolus maximum
in one hour causes the following message to be displayed:
Press SEL and ACT, and then place the PPC near the Pump and allow the com-
munication to complete. The patient may exceed the pre-programmed bolus limit
by programming another bolus within 10 minutes.
PPC Display Type of Alarm
PUMP
SUSPENDED
2
PPC Display Type of Alarm
AUTO OFF PUMP
PUMP SUSPENDED
1
PPC Display Type of Alarm
HOURLY MAX
EXCEEDED
1