User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
System alarms and messages
94
Battery replacement
If the PPC main battery (AA 1.5V alkaline) has been replaced or the PPC recog-
nizes the “PUMP STATUS” needs to be checked, the following message is dis-
played:
The user should reposition the PPC near the Pump, and then press SEL and ACT.
The PPC will communicate with the Pump.
NOTE: If while programming the PPC, the screen goes blank, the
PPC beeps six times and then the “CHECK PUMP
STATUS” message appears, the battery needs to be
replaced. (See Chapter 3 for instructions, Install/Replace
the Main Battery.)
Initialize to factory defaults
When the Pump is reinitialized to its factory default settings, the following mes-
sage is displayed:
Pump stopped
If the Pump is intentionally stopped, the following message is displayed:
To restart the Pump, press SEL and ACT, and then place the PPC near the Pump
and allow the communication to complete. The Pump can only be stopped in the
Supervisor Mode.
PPC Display Type of Alarm
CHECK
PUMP STATUS
1
PPC Display Type of Alarm
PUMP
RESET
1
PPC Display Type of Alarm
PUMP
STOPPED
2