User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
System alarms and messages
92
PPC alarms
The PPC offers a choice of two alarms, audible and vibrate. In addition, a screen
message appears indicating the type of alarm condition that occurred.
Low reservoir
When the Pump calculates that less than 800 units (2ml) of insulin remains in its
reservoir, the following display will appear:
Clear this message by pressing SEL and ACT. Users should be instructed to
report this alarm to their physician as soon as possible and schedule an appoint-
ment for a Pump refill.
Empty reservoir
When the Pump calculates that less that 400 units (1ml) of insulin remain in its
reservoir, the following display will appear:
Clear this message by pressing SEL and ACT, and then continue programming.
Users should be instructed to report this alarm to their physician as soon as possi-
ble and schedule an appointment for a Pump refill. It is important not to allow the
Pump to deplete its insulin supply as this may result in catheter blockage.
Telemetry communication error
If programming is interrupted after partial transmission of a command, the PPC
will display the following message on the display screen:
Reposition the PPC near the Pump, and then press SEL and ACT. The PPC will
attempt to resume communication with the Pump.
PPC Display Type of Alarm Pump Alarm
LOW
RESERVOIR
1 In 24 hours if no PPC
communication
PPC Display Type of Alarm Pump Alarm
EMPTY
RESERVOIR
1 In 24 hours if no PPC
communication
PPC Display Type of Alarm
TELEMETRY COMM ERROR3 1