User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
System alarms and messages
91
System error
The Implantable Insulin Pump has a sophisticated self-monitoring system that
periodically checks for circuit faults. If a fault should occur in the Pump electron-
ics, insulin delivery will stop. Conventional insulin therapy must be initiated
immediately. Notify Medtronic MiniMed immediately.
Pump self test fail
If during a “SELF TEST” the Pump presents a malfunction, the insulin delivery
will stop. Clear the message by pressing SEL then ACT. Notify Medtronic Min-
iMed immediately.
PPC low battery
If the PPC main battery (AA 1.5 volt alkaline) energy is low, the following alarm
display appears each time a new function is programmed:
Clear this message by pressing SEL and then ACT. While the battery should
have sufficient energy for a few additional programming commands, the battery
should be changed as soon as possible. For instructions on changing the battery,
(see Chapter 3, Personal Pump Communicator).
NOTE: If while programming the PPC, the screen goes blank, the
PPC beeps six times and then the “CHECK PUMP
STATUS” message appears, the battery needs to be
replaced. (See Chapter 3 for instructions, Install/Replace
the Main Battery).
PPC Display Type of Alarm Pump Alarm
PUMP STOPPED
1 or 2 or 3 or 4 or 5 or 6
1 In 5 minutes if no PPC
communication
PPC Display Type of Alarm
PUMP
SELF TEST FAIL
1
PPC Display Type of Alarm
PPC
LOW BATTERY
1