Alarms and Messages
50
Depleted pump battery
When there is no longer sufficient battery energy to power the Pump, the
Pump Low Battery Alarm will cease. This means that insulin delivery has
stopped. You must switch to another type of insulin therapy.
System error
The Implantable Insulin Pump has a sophisticated self-monitoring system
that continuously checks for circuit faults. If the Pump detects a circuit
error, it will beep and then automatically shut itself off. When this
occurs, you must switch to your alternative insulin therapy estab-
lished between you and your doctor. Notify your doctor immediately of
the System Error.
Pump self test fail
During a SELF TEST, the Pump finds a malfunction. The Pump will
stop delivering insulin. Clear the message by pressing SEL then ACT.
You must switch to your alternative insulin therapy established
between you and your doctor. Notify your doctor immediately of the
Self Test Error.
PPC Display Type of Alarm Pump Alarm
PUMP STOPPED
1or2or3or4or5or6
1 In 5 minutes if no PPC
communication
PPC Display Type of Alarm
PUMP
SELF TEST FAIL
1