User Manual

Table Of Contents
Programming Your Pump
37
Alarms
There are three PPC Alarm options, two audible tones (Low/High) and a
vibrate mode.
ALARMS will alert you if either the PPC or Pump recognizes a prob-
lem. Your doctor or nurse will make sure the ALARM FEEDBACK
screen is in the ON position.
Vibrate only for alarms, all programming feedback from PPC is done with
beeps.
NOTE: If you want the PPC to vibrate instead of beep, the
PPC battery will only last 6 weeks instead of 8
weeks, under conditions of normal use.
1. Press ACT to enter the ALARMS
menu.
2. Press the
and
buttons to select the
type of alarm you want, then press ACT.
3. The screen will now display SET PPC
FEEDBACK. This setting should
always be ON.PressACT.
4. Place the PPC near the Pump. When the
communication is completed, the PPC
will change to SELF TEST and then
timeout to the Time / Day screen.
ALARMS
SET
PPC
ALARM TYPE
LOW/HIGH/VIBRATE
SET
ALARM
FEEDBACK
ON/OFF
PPC
COMMUNICATING