User Manual

Table Of Contents
Programming Your Pump
36
Auto off
AUTO OFF sets an alarm on your PPC, to remind you if you have not
programmed your Pump recently. The AUTO OFF alarm can be set
between 1 and 16 hours. To turn off "Auto Off" program set the time to
dashes.
1. From the SETUP PUMP screen, press
ACT then SEL.
2. The screen will display AUTO OFF
with flashing dashes. Press ACT.
3. Use the
and
buttons to select the
number of hours before an Auto Off
alarm occurs.
4. Press ACT. In this example, a time dura-
tion of 10 hours was selected. The PPC
will alarm if you did not program your Pump during the past 10
hours.
5. Place the PPC near the Pump. The PPC
will beep once when the communication
is completed.
AUTO OFF
--
HRS
AUTO OFF
10
HRS
SET
PPC
COMMUNICATING