System Alarms and Messages
122
PPC needs servicing
When the PPC internal battery (lithium) energy becomes low, the follow-
ing message will be displayed:
Clear this message by pressing SEL and ACT. Users should be instructed
to report this alarm to their physician as soon as possible and have their
PPC replaced. If the PPC loses all power, it may also lose information
stored in memory.
Low reservoir
When the Pump calculates that less than 800 units (2 ml) of insulin
remains in its reservoir, the following display will appear:
Clear this message by pressing SEL and ACT. Users should be instructed
to report this alarm to their physician as soon as possible and schedule an
appointment for a Pump refill.
Empty reservoir
When the Pump calculates that less that 400 units (1ml) of insulin remains
in its reservoir, the following display will appear :
PPC Display Type of Alarm
PPC
NEEDS SERVICING
1
PPC Display Type of Alarm Pump Alarm
LOW
RESERVOIR
1 In 24 hours if no
PPC
communication
PPC Display Type of Alarm Pump Alarm
EMPTY
RESERVOIR
1 In 24 hours if no
PPC
communication
Dmp9196021-011_c.book Page 122 Wednesday, April 3, 2002 5:06 PM