Technical Specs Part 3

133
CHAPTER 11 Medtronic MiniMed 2007C
Implantable Insulin Pump
System
Implantable Insulin Pump MMT-2007C
Diameter
Thickness
Reservoir Volume
8.1 cm (3.2 inches)
2.0 cm (0.8 inches)
13 ml to 15 ml
Weight - Empty 131 gm (4.6 ounces)
Insulin - Concentration Aventis HOE 21 PH U-400
Stroke Volume 0.42 to 0.58 ml per stroke
0.17 to 0.23 units per stroke
Basal Rate 0.2 to 35.0 units per hour (U/h)
Basal Patterns 3 patterns of up to 48 basal rates each.
Meal Bolus 0.2 to 25.0 units
Bolus Duration Immediate, Square Wave (30 minutes
to 4 hours), or both together. Audio
Bolus
Temporary Basal Rate 0.2 to 35.0 units per hour
30 minute increment duration
30 minutes up to 24 hours delay
Diagnostic Rate 10 to 150 U/h
Power Supply Lithium - Carbon Monofluoride Bat-
tery
Battery life See Figure 1
Dmp9196021-011_c.book Page 133 Wednesday, April 3, 2002 5:06 PM