Technical Specs Part 3
Table Of Contents
- CHAPTER 9 System Alarms and Messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC alarms
- PPC low battery
- PPC depleted battery
- PPC needs servicing
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System Under-delivery
- CHAPTER 11 Medtronic MiniMed 2007C Implantable Insulin Pump System
System Alarms and Messages
126
Initialize to factory defaults
When the Pump is reinitialized to its factory default settings, the follow-
ing message is displayed:
Pump stopped
If the Pump is intentionally stopped, the following message is displayed:
To restart the Pump, press SEL and ACT, and then place the PPC near
the Pump and allow the communication to complete. The Pump can only
be stopped in the Supervisor Mode.
Pump suspended
If the Pump operation has been suspended, the following message is dis-
played:
During “SUSPEND PUMP”, the Pump will deliver a basal rate of
0.2 U/h. To restart insulin delivery programming, press SEL and ACT,
then place the PPC near the Pump and allow the communication to com-
plete.
PPC Display Type of Alarm
PUMP
RESET
1
PPC Display Type of Alarm
PUMP
STOPPED
2
PPC Display Type of Alarm
PUMP
SUSPENDED
2