Manual Part 2

90
Figure 9-3: Tubing and Retainer Removed.
Step 4 Inspect the outlet of the Pump for debris. Debris, espe-
cially fibers, left in the fluid path will compromise the
function of the Catheter check valve. Rinse the outlet
tube of the Pump with water from a syringe, if necessary.
Avoid wiping the Pump with gauze or drapes that will
leave fibrous debris.
Use caution when removing debris from
the outlet of the Pump. There is a precision
rubber seal on the outlet that will be dam-
aged by handling with surgical instru-
ments.
Step 5 Pull the locking bar out and gently push the Catheter
straight down onto the Pump outlet. Never twist the
connector.
!
WARNING
Dmp9196021-011_c.book Page 90 Wednesday, April 3, 2002 5:04 PM