Manual Part 2

Pump Refill Procedure
98
Supplies and solutions
The refill kit (MMT-4105) is intended exclusively for use with the
Medtronic MiniMed 2007C Implantable Insulin Pump System. The refill
kit consists of a specially designed syringe and stopcock for removing flu-
ids from and placing fluids into the Pump. Use only refill needles (MMT-
4102) available from MiniMed to perform all Pump refill procedures.
This 22 gauge needle, featuring a rounded, lubricated tip and a beveled
side hole, has been specially developed to mate with the Pump fill port.
Use of other needles may result in damage
to the Pump septum in the fill port, and
may allow body fluids to enter the Pump.
Below is a list of the supplies and solutions necessary to perform the refill
procedure:
Supplies:
Steri-Strips
®
and Markers
The Patients PPC
Glucose Monitoring Equipment
•Drapes
One (1) Scale (0.01 gram resolution)
One (1) 250 ml Sterile Beaker
One (1) Port Locating Template - MMT-4106
Two (2) Refill Kits - MMT-4105
Three (3) MiniMed MMT-4102 Refill Needles (Extra needles
should be available for use.)
Four (4) Sharp 18 Gauge Regular Bevel Needles
Solutions:
One (1) 10 ml Vial, Aventis Rinse Buffer Solution
Two (2) 10 ml Vials, Aventis HOE 21 PH, U-400 Insulin
!
WARNING
Dmp9196021-011_c.book Page 98 Wednesday, April 3, 2002 5:04 PM