Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Setup II
73
PUMP INITIALIZED The PPC has been “married” to a Pump . Press SEL
and ACT.
PRIMING XX U A Priming Bolus has been programmed and is being
delivered.
PUMP RESET The Pump was reprogrammed to the preset factory val-
ues.
PUMP SELF TEST FAIL A Pump malfunction was detected during a Self Test.
Notify your physician immediately.
PUMP STOPPED The Pump is in stop mode.
PUMP STOPPED
1 or 2 or 3 or 4or 5 or 6
When the Pump recognizes a system malfunction, it
automatically stops and insulin delivery ceases. Notify
your physician immediately.
PUMP SUSPENDED The Pump is in suspend mode
PUMP ERROR
0 or 1
User attempts to initialize the PPC to a Pump that is not
compatible with it. Clear by pressing SEL and ACT.
Verify the personal ID of the Pump responding corre-
sponds to the personal ID entered into the PPC.
PUMP ERROR
40 or 41
The Pump has invalid data. The PPC will not initialize
to Pump.
TELEMETRY COMM
ERROR
3
The PPC and Pump are not communicating. Reposi-
tion the PPC over the Pump, then press SEL and ACT.
If the error message persists, contact MiniMed.
TELEMETRY COMM
ERROR
20
If the response is “NO” to the “INITIALIZE PPC TO
PUMP” until this screen appears. Move the PPC away
from any other Pump and perform the request again.
Display Screen Message Message Meaning or Action Required