Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Personal Pump Communicator (PPC)
72
Exit supervisor
To exit the Supervisor Programming Menu, press SEL until the “EXIT
SUPERVISOR” screen appears, then press ACT.
Personal Pump communicator messages
Display Screen Message Message Meaning or Action Required
AUTO OFF PUMP PUMP
SUSPENDED
Auto Off time interval has elapsed. Pump operation is
suspended.
BOLUS 0.0 u A bolus has been programmed and is being delivered.
CHECK PUMP
STATUS
The PPC battery has been replaced. The PPC needs to
check the Pump status.
LOW RESERVOIR The PPC has recognized the Pump has 800 units (2 ml)
or less insulin remaining in its reservoir. Schedule a
Pump refill as soon as possible. Allowing the reservoir
to completely empty may damage the Pump.
EMPTY RESERVOIR The PPC has recognized the Pump has 400 units (1 ml)
or less insulin remaining in its reservoir. Schedule a
Pump refill as soon as possible. Allowing the reservoir
to completely empty may damage the Pump.
HOURLY MAX
EXCEEDED
You attempt to deliver more than 2.5 times the bolus
maximum in one hour. To clear the message, press
SEL and ACT. You may exceed this limit by program-
ming another bolus within 10 minutes.
PPC NEEDS
SERVICING
The internal PPC backup battery is depleted. The PPC
can be programmed. Replace the PPC as soon as possi-
ble.
PPC LOW BATTERY The internal PPC backup battery is low. The PPC can
be programmed. Replace the PPC as soon as possible.
PPC DEPLETED
BATTERY
The PPC cannot be programmed. Replace the PPC AA
1.5 volt alkaline battery.
PPC NOT INITIALIZED The PPC has not been initialized to a Pump.
PPC LOW BATTERY The PPC can be programmed. Replace the PPC AA 1.5
volt alkaline battery.
PUMP LOW
BATTERY
The Pump battery energy is low but still functioning.
Schedule a replacement as soon as possible.