Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Personal Pump Communicator (PPC)
62
Set an audio bolus from the main menu bolus screen
1. From the Time/Date screen press the ▲
▲▲
▲
button. The PPC beeps either one or two
times, depending upon the audio bolus
step amount set. The user counts the
number of beeps to determine how much
insulin was programmed.
2. In this example the PPC was programmed
for a 0.8U increment, and the PPC beeps
two times for the 0.8U bolus.
Press ACT and the audio sequence
repeats to confirm the bolus amount.
Press ACT twice and the Pump will be
programmed. The PPC beeps once when the programming is com-
pleted. Place the PPC on the Pump to complete the communication
process.
NOTE: While an audio bolus is being delivered, the
▲
▲▲
▲
button will not function.
Variable bolus
To access the Variable Bolus feature enter “SETUP II” and press SEL
one time.
The screen will display “VARIABLE
BOLUS”. Press ACT.
The screen will now display “SET VARI-
ABLE BOLUS” and “ON or OFF” will be
flashing. Use the ▲
▲▲
▲
and ▼
▼▼
▼
buttons to select
“ON” or “OFF”, then Press ACT.
SET BOLUS
0.8U
IMM
PPC
COMMUNICATING
VARIABLE
BOLUS
ON
SET
VARIABLE
BOLUS
Dmp9196021-011_c.book Page 62 Thursday, April 4, 2002 8:15 AM