Technical Specs Part 3
Table Of Contents
- CHAPTER 9 System Alarms and Messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC alarms
- PPC low battery
- PPC depleted battery
- PPC needs servicing
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System Under-delivery
- CHAPTER 11 Medtronic MiniMed 2007C Implantable Insulin Pump System
Pump alarms
119
The Pump Low Battery alarm can be cleared and the Pump will continue
to operate normally. However, users should be instructed to report the
alarm immediately to their physician. A Pump replacement or resumption
of conventional insulin therapy should then be scheduled.
Depleted pump battery
When there is no longer sufficient battery energy to power the Pump, the
Pump Low Battery Alarm will cease and insulin delivery will stop. Alter-
nate insulin therapy must be initiated.
System error
The Implantable Insulin Pump has a sophisticated self-monitoring system
that periodically checks for circuit faults. If a fault should occur in the
Pump electronics, insulin delivery will stop. Conventional insulin therapy
must be initiated immediately. Using the PPC Supervisor Mode, the phy-
sician should use the “DOWNLOAD SOFTWARE” feature to reprogram
the Pump from software in the PPC memory. Notify MiniMed immedi-
ately.
PPC Display Type of Alarm Pump Alarm
PUMP
LOW BATTERY
1 In 24 hours if no PPC
communication
PPC Display Type of Alarm Pump Alarm
PUMP STOPPED
1 or 2 or 3 or 4 or 5 or 6
1 In 5 minutes if no PPC
communication
Dmp9196021-011_c.book Page 119 Wednesday, April 3, 2002 5:06 PM