Pump alarms
127
Auto off
If the “AUTO OFF” time interval elapses in the Pump, and the PPC rec-
ognizes this condition, the following message is displayed:
Press SEL and ACT, and then place the PPC near the Pump and allow the
communication to complete. The PPC then communicates with the Pump
to reset the “AUTO OFF” duration.
If five more minutes elapse, the Pump will initiate the internal alarm
sequence of a beep every 15 seconds for 10 minutes, then double-beeps
every 15 seconds for 10 minutes, then repeating the pattern. The alarm is
cleared by pressing SEL and ACT.
Hourly maximum exceeded
Attempting to deliver more than 2.5 times the pre-programmed bolus
maximum in one hour, causes the following message to be displayed:
Press SEL and ACT, and then place the PPC near the Pump and allow the
communication to complete. The patient may exceed the pre-programmed
bolus limit by programming another bolus within 10 minutes.
PPC Display Type of Alarm
AUTO OFF PUMP
PUMP SUSPENDED
1
PPC Display Type of Alarm
HOURLY MAX
EXCEEDED
1
Dmp9196021-011_c.book Page 127 Wednesday, April 3, 2002 5:06 PM