Product Info Part 4

Table Of Contents
193
1. Maintain syringe 6 with approximately 10 ml of residual INSULIN in
the Pump fill port. Verify that the stopcock on syringe 6 is open.
2. Prime the syringe 5 needle. Close the stopcock.
3. Obtain a vacuum in syringe 5 by pulling back on the plunger until it
locks. Press the lock into the plunger groove to be sure it is firmly
secured.
4. Enter the side port with syringe 5 - stopcock closed.
5. Open the stopcock on syringe 5.
6. Program and deliver a 4 unit bolus to open the valve in the Pump
mechanism.
7. Observe syringe 6. If the level of INSULIN approaches 2 ml, close
the stopcock to prevent air from entering the Pump. Observe syringe
5. At least 1 ml of INSULIN must pass into syringe 5 to be sure that
all of the RINSE BUFFER is removed from the side port. If less than
1 ml enters, repeat the bolus from Step 6, some Pumps may require 3
successive boluses to obtain 1 ml.
8. Close both stopcocks. Remove syringe 6 from the Pump. Set aside
the syringe to be weighed.
9. With the stopcock closed, remove syringe 5 from the side port.
Syringe 5 will have a vacuum inside from the previous step. Enter the
Pump fill port with syringe 5 and open the stopcock to remove 2 ml.
This will restore the negative pressure safety feature of the Pump.
10. Close the stopcock, remove syringe 5 from the Pump, and set it aside
to be weighed.
Dmp9196021-011_c.book Page 193 Wednesday, April 3, 2002 5:07 PM