Product Info Part 4
Table Of Contents
- APPENDIX D Precautions and General Procedures
- Special note and precautions
- General procedures
- Locating the pump fill port and the side port
- Venting the Minimed refill syringe (optional)
- APPENDIX E Pump Rinse Procedure
- Preparing for the procedure
- Remove insulin from the Pump and fill with NaOH
- Equilibrate and pull NaOH through system
- APPENDIX F Side Port Catheter Flush Procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke Volume Measurement
- Supplies and Solutions
- APPENDIX H Pressure Measurement Using The Side Port Catheter
- Supplies and solutions
- Measuring pressure in the side port
169
Equilibrate and pull rinse buffer through system
The reservoir pressure in the Pump is equilibrated with outside ambient
pressure and RINSE BUFFER is pulled through the fluid system. This
procedure uses syringes 2 and 3:
• Syringe 2 with RINSE BUFFER. Expel all RINSE BUFFER
except 5 mL. Change the refill needle if desired.
• Syringe 3 with residual RINSE BUFFER from the previous sec-
tion.
Syringe 3: Rinse Buffer
1. Maintain syringe 3 with approximately 10 mL of RINSE BUFFER in
the Pump fill port. Verify that the stopcock on syringe 3 is open.
Syringe 2: Rinse Buffer (residual)
2. Prime the needle. Close the stopcock.
3. Obtain a vacuum by pulling back on the plunger until it locks. Press
the lock into the plunger groove to be sure it is firmly secured.
4. Enter the side port with the stopcock closed.
5. Open the stopcock.
6. Program and deliver a 4 unit bolus to open the valve in the Pump
mechanism.
7. Observe syringe 3. When the level of the RINSE BUFFER
approaches 2 mL, close the stopcock to prevent air from entering the
Pump. Observe syringe 2. When at least 1 mL of RINSE BUFFER
has entered syringe 2 close both stopcocks and remove the syringes.
If less than 1 ml enters, repeat the bolus from Step 6, some Pumps
may require 3 successive boluses to obtain 1 ml. Discard syringe 2
and set aside syringe 3 to be used later to remove the RINSE
BUFFER.
Dmp9196021-011_c.book Page 169 Wednesday, April 3, 2002 5:07 PM