PUMP INITIALIZED The PPC has been “married” to a Pump . Press SEL
and ACT.
PRIMING XX U A Priming Bolus has been programmed and is being
delivered.
PUMP RESET The Pump was reprogrammed to the preset factory val-
ues.
PUMP SELF TEST FAIL A Pump malfunction was detected during a Self Test.
Notify your physician immediately.
PUMP STOPPED The Pump is in stop mode.
PUMP STOPPED
1 or 2 or 3 or 4or 5 or 6
When the Pump recognizes a system malfunction, it
automatically stops and insulin delivery ceases. Notify
your physician immediately.
PUMP SUSPENDED The Pump is in suspend mode
PUMP ERROR
0 or 1
User attempts to initialize the PPC to a Pump that is not
compatible with it. Clear by pressing SEL and ACT.
Verify the personal ID of the Pump responding corre-
sponds to the personal ID entered into the PPC.
PUMP ERROR
40 or 41
The Pump has invalid data. The PPC will not initialize
to Pump.
TELEMETRY COMM
ERROR
3
The PPC and Pump are not communicating. Reposi-
tion the PPC over the Pump, then press SEL and ACT.
If the error message persists, contact MiniMed.
TELEMETRY COMM
ERROR
20
If the response is “NO” to the “INITIALIZE PPC TO
PUMP” until this screen appears. Move the PPC away
from any other Pump and perform the request again.
Display Screen Message Message Meaning or Action Required