Installation Instructions Part 1

Table Of Contents
Personal Pump Communicator (PPC)
72
Exit supervisor
To exit the Supervisor Programming Menu, press SEL until the “EXIT
SUPERVISOR” screen appears, then press ACT.
Personal Pump communicator messages
Display Screen Message Message Meaning or Action Required
AUTO OFF PUMP PUMP
SUSPENDED
Auto Off time interval has elapsed. Pump operation is
suspended.
BOLUS 0.0 u A bolus has been programmed and is being delivered.
CHECK PUMP
STATUS
The PPC battery has been replaced. The PPC needs to
check the Pump status.
LOW RESERVOIR The PPC has recognized the Pump has 800 units (2 ml)
or less insulin remaining in its reservoir. Schedule a
Pump refill as soon as possible. Allowing the reservoir
to completely empty may damage the Pump.
EMPTY RESERVOIR The PPC has recognized the Pump has 400 units (1 ml)
or less insulin remaining in its reservoir. Schedule a
Pump refill as soon as possible. Allowing the reservoir
to completely empty may damage the Pump.
HOURLY MAX
EXCEEDED
You attempt to deliver more than 2.5 times the bolus
maximum in one hour. To clear the message, press
SEL and ACT. You may exceed this limit by program-
ming another bolus within 10 minutes.
PPC NEEDS
SERVICING
The internal PPC backup battery is depleted. The PPC
can be programmed. Replace the PPC as soon as possi-
ble.
PPC LOW BATTERY The internal PPC backup battery is low. The PPC can
be programmed. Replace the PPC as soon as possible.
PPC DEPLETED
BATTERY
The PPC cannot be programmed. Replace the PPC AA
1.5 volt alkaline battery.
PPC NOT INITIALIZED The PPC has not been initialized to a Pump.
PPC LOW BATTERY The PPC can be programmed. Replace the PPC AA 1.5
volt alkaline battery.
PUMP LOW
BATTERY
The Pump battery energy is low but still functioning.
Schedule a replacement as soon as possible.