Initialize to factory defaults
This feature resets all programmed parameters in the Pump to preset fac-
tory default values. To access this feature, the Pump must be in the
“PUMP STOPPED” mode.
1. From the “PUMP STOPPED” screen,
press SEL. The screen will begin flash-
ing.
2. Press the
and
buttons at the same
time until the “ENTER SUPERVISOR
PASSWORD” screen appears. Reenter
the Supervisor Password, pressing ACT
after each letter or number is entered.
The factory preset password is: YIQ8.
3. When the Supervisor Password is suc-
cessfully entered, the screen will change
to “PUMP REFILL”. Press SEL until the
“INITIALIZE TO FACTORY
DEFAULTS” screen appears.
4. Press ACT and the word “CONFIRM”
will appear. Press ACT again to activate.
5. Place the PPC over the Pump and com-
plete the communication process.
6. The PPC will now read “PUMP SUS-
PENDED.” Press SEL then ACT.
08:13 oct 12
PUMP STOPPED
ENTER
PASSWORD
SUPERVISOR
0000
INITIALIZE
DEFAULTS
TO FACTORY
INITIALIZE
DEFAULTS
TO FACTORY
CONFIRM
PPC
COMMUNICATING
08:13 OCT 02
PUMP SUSPENDED