Installation Instructions Part 1

Table Of Contents
Setup II
65
Priming
A priming bolus is not intended as a therapeu-
tic bolus. It should be only used when the
Pump is not implanted.
The Priming Bolus is used during the preparation of the Pump for implan-
tation. This feature “Primes” the Pump piston chamber, outlet port and
Catheter. The Pump will pulse approximately 500 times during this func-
tion. To access this feature, the Pump must be in “PUMP STOPPED”
mode.
1. From the “PUMP STOPPED” screen,
press SEL. The screen will begin flash-
ing.
2. Press the
and
buttons at the same
time until the “ENTER SUPERVISOR
PASSWORD” screen appears.
3. Re-enter the supervisor password, press-
ing ACT after each letter or number is
entered. The factory preset password is:
YIQ8. If the password has been changed,
enter the new password.
When the Super-
visor Password is successfully entered,
the screen will change to “PUMP REFILL”. Press SEL until the
“PRIMING” screen appears.
4. Press ACT and “NO” appears flashing.
Press the
and
buttons to change to
“YES”, then press ACT.
!
WARNING
08:13 Oct 12
PUMP STOPPED
ENTER
SUPERVISOR
PASSWORD
YIQ8
PRIMING
PRIMING
PRIMING NO
99.8u
Dmp9196021-011_c.book Page 65 Thursday, April 4, 2002 8:15 AM