Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Setup II
65
Priming
A priming bolus is not intended as a therapeu-
tic bolus. It should be only used when the
Pump is not implanted.
The Priming Bolus is used during the preparation of the Pump for implan-
tation. This feature “Primes” the Pump piston chamber, outlet port and
Catheter. The Pump will pulse approximately 500 times during this func-
tion. To access this feature, the Pump must be in “PUMP STOPPED”
mode.
1. From the “PUMP STOPPED” screen,
press SEL. The screen will begin flash-
ing.
2. Press the ▲
▲▲
▲ and ▼
▼▼
▼ buttons at the same
time until the “ENTER SUPERVISOR
PASSWORD” screen appears.
3. Re-enter the supervisor password, press-
ing ACT after each letter or number is
entered. The factory preset password is:
YIQ8. If the password has been changed,
enter the new password.
When the Super-
visor Password is successfully entered,
the screen will change to “PUMP REFILL”. Press SEL until the
“PRIMING” screen appears.
4. Press ACT and “NO” appears flashing.
Press the ▲
▲▲
▲ and ▼
▼▼
▼ buttons to change to
“YES”, then press ACT.
!
WARNING
08:13 Oct 12
PUMP STOPPED
ENTER
SUPERVISOR
PASSWORD
YIQ8
PRIMING
PRIMING
PRIMING NO
99.8u