Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Setup II
61
Setup II
Audio bolus
The Audio Bolus feature allows the user to deliver a bolus without look-
ing at the PPC screen. There are two Audio Bolus increments, 0.4U and
0.8U. With each press of the ▲
▲▲
▲ button, the PPC will beep one time (0.4U
setting) or two times (0.8U setting), depending on the delivery amount
selected per button press.
Activating the audio bolus feature
1. From the “SETUP II” screen, press ACT.
The “AUDIO BOLUS’ screen is shown.
2. Press ACT, and the screen changes to
“SET AUDIO BOLUS” with “OFF/ON”
flashing. Use the ▲
▲▲
▲ and ▼
▼▼
▼ buttons to
choose “ON” to activate the Audio Bolus
feature or “OFF” to disable it. Then press
ACT.
3. If “ON” was chosen, the “SET STEP
AUDIO BOLUS” appears with a step
value of 0.4U or 0.8U flashing on the
screen. Use the ▲
▲▲
▲ and ▼
▼▼
▼ buttons to
choose between a step rate of 0.4U or
0.8U. Press ACT. The Audio Bolus fea-
ture is now activated.
NOTE: If “OFF” is chosen, pressing the
▲
▲▲
▲
button will
have no effect.
AUDIO BOLUS
ON 0.4u
SET
ON
AUDIO BOLUS
0.4u
SET STEP
AUDIO BOLUS
0.4u