Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Personal Pump Communicator (PPC)
54
Stop a temporary basal rate
1. From the Time/Date screen press SEL
until the “TEMP BASAL” screen is dis-
played. Press ACT and the “SET DURA-
TION” appears, with the time duration
flashing. Press ▼
▼▼
▼ once until it resets to
dashes. Then press ACT.
2. Place the PPC near the Pump and com-
plete the communication process.
3. When the communication is complete, the
PPC will briefly return to the “TEMP
BASAL” screen with the amount dashes
flashing. Allow the PPC to return to the
Time/Date screen.
Personal events
This feature allows the user to enter event codes into the PPC memory,
and record the time and date of entry. Preset event codes are: 1 = meal, 2
= snack, 3 = sick and 4 = exercise. In addition, other event codes A, B and
C can be entered to record other important events. These other event
codes should be documented prior to their use.
To access the “EVENTS” screen in the main menu, “ON” must be acti-
vated in the “SETUP II” menu, “PERSONAL EVENTS” screen.
1. From the Time/Date screen, press SEL
until “SETUP PUMP” is displayed, then
press ACT.
Press SEL until “SETUP II” is displayed,
then press ACT.
Press SEL until “PERSONAL EVENTS”
is screen is displayed.
TEMP BASAL
--- --- 1.5u/h
SET DURATION
PPC
COMMUNICATING
TEMP BASAL
--- --- --- --- u/h
SET AMOUNT
OFF
PERSONAL
EVENTS