Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Personal Pump Communicator (PPC)
34
Set the time and date
The time and date settings must be correct to ensure accurate calculation
of insulin delivery, daily totals, and the proper display of insulin activity
history.
1. Press SEL until the “SETUP PUMP”
screen is displayed, then press ACT two
times. The first two digits of the time
(hours) will be flashing. Use the ▲
▲▲
▲ and ▼
▼▼
▼
buttons to select the correct hour, then
press ACT. The last two digits of the time
(minutes) will be flashing. Use the ▲
▲▲
▲ and ▼
▼▼
▼ buttons to select the cor-
rect minute, then press ACT. Repeat the programming process to
enter information for the year, month and day.
2. After completing the programming pro-
cess, quickly place the PPC near the
Pump. The PPC will display “PPC
COMMUNICATING” while transfer-
ring the time and date information to the
Pump. The PPC will then move to the
next screen, “AUTO-OFF.” Skip the “AUTO-OFF” screen by press-
ing SEL once to reach the next screen, “ALARMS”.
08:32 Jan 02
SET 2000
TIME-DATE
PPC
COMMUNICATING