Initialize the PPC
When the healthcare professional receives a new PPC it must be “mar-
ried” to the Pump. Following are the basic steps used to initialize a Pump
System the day prior to implant:
1. The PPC is delivered without a battery in
place. After installing a new battery, the
PPC will beep six times and the screen
will identify the PPC software used (see
Chapter 1, Install/Replace the Main
Battery).
2. The screen now changes to, “PPC NOT
INITIALIZED”, and the PPC will beep six
times once every minute until the initial-
ization process is started. Press SEL and
then ACT, then quickly place the PPC
over the Pump.
3. When a communication link has been
established, the screen will read, “PPC
COMMUNICATING”, and then will
change to the next screen.
4. NO is blinking. Check to make sure the
serial number displayed on the screen
matches the Pump serial number. Press
either
or
once to change “NO” to
“YES” and then press ACT. Place the
PPC over the Pump.
5. The screen again reads “PPC COMMUNI-
CATING”, and the PPC will beep 3 times
at the end of the programming sequence.
During this process, the PPC receives all
of the factory preset values contained in
the Pump memory.
PUMP
PPC
4 162 100
__ __ __
PPC
NOT INITIALIZED
PPC
COMMUNICATING
INITIALIZING "NO"
PPC to PUMP
..................................
........................
47568
COMMUNICATING
PPC