Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Introduction
29
Install/Replace the main battery
The battery used to power the PPC is a 1.5v AA alkaline. The life of the
battery is approximately 8 weeks during normal usage conditions. If the
vibrator mode is selected, the battery life is approximately 6 weeks.
• Locate the battery door on the back of the PPC.
• Slide the locking bar to the left.
• Push the middle part of the PPC box (under the battery door) and
lift by gently pulling up the battery door to unlatch.
• Remove the old battery, noting the polarity. The screen will be
blank.
• Position the new battery so the + and - markings on the battery
match the polarity diagram in the battery compartment.
• Close the battery door.
• Slide the locking bar to the right.
• The PPC screen will reappear within 30 seconds:
1. The PPC will beep 6 times, and after a
few seconds, the screen will display
“CHECK PUMP STATUS”.
2. Press SEL then ACT, and place the PPC
near the Pump.
3. Wait a few seconds for the communica-
tion to complete.
PPC 4 102 100
PUMP __ __ __
08:26 JAN O2
CHECK
PUMP STATUS
PPC
COMMUNICATING
Dmp9196021-011_c.book Page 29 Thursday, April 4, 2002 8:15 AM