User Manual

Table Of Contents
Alarms and Messages
48
The PPC alarms
The PPC has three types of alarms, audible or vibrate alarms and visual
alarms. This chapter will describe, for each alarm condition, which screen
message appears.
Some alarms can be cleared by pressing SEL then ACT. The difference
between these alarms will be explained later in this chapter.
Two audible or vibrate alarm types are used:
Alarm type 1
If the PPC is set to vibrate, the vibrator will be turned on for 3 seconds
every minute until the condition is cleared.
If the PPC is set to audible, the PPC will beep 6 times every minute for
30 minutes. If the alarm is not cleared in 30 minutes, the PPC will beep 6
alternating tones. The PPC will continue to do so every minute until the
condition is cleared.
Alarm type 2
If the PPC is set to vibrate, the vibrator will be turned on for 3 seconds
every minute while the condition exists.
If the PPC is set to audible, the PPC 6 will beep 3 times every 30 min-
utes while the condition exists.