User's Manual

Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The MiniMed 640G insulin pump is intended for use in the electromagnetic
environment specified below. The customer or user of the MiniMed insulin pump
should assure that it is used in such an electromagnetic environment.
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to
800 MHz
10 V/m
800 MHz to
2.5 GHz
10 V/m
80 MHz to
800 MHz
10 V/m
800 MHz to 6
GHz
Recommended separation distance
d = [12/E
1
]P
80 MHz to 800 MHz
d = [23/E
1
]P
800 MHz to 6 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
a
, should
be less than the compliance level in
each frequency range
b
.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption, and reflection from structures, objects and people.
Note: The table is per IEC (EN) 60601-1-2 Edition 3.
product specifications and safety information
Product specifications and safety information 265
MP6025958-022 / A
RELEASED