User's Manual
Percent temp basal
The default value is 100 percent of basal programming.
Range 0 to 200%
Default 100% of basal programming
Increment 5%
Program safety checks
A single fault condition will cause the pump to suspend insulin delivery. Maximum
infusion with a single fault condition is 0.2 units.
Pump dimensions
The MMT-1512 pump dimensions in inches are approximately
2.1 width x 3.34 length x 0.96 depth.
The MMT-1512 pump dimensions in centimeters are approximately 5.3 width x 8.5
length x 2.44 depth.
The MMT-1712 pump dimensions in inches are approximately
2.1 width x 3.78 length x 0.96 depth.
The MMT-1712 pump dimensions in centimeters are approximately 5.3 width x 9.6
length x 2.44 depth.
Pump memory
User settings and pump history are stored in non-volatile memory which will retain
data. The memory size will hold 90 days of pump history before it becomes full
and has to be written over. This means that at any time the user can review a
maximum of 90 days of history.
Pump weight
The mass of the MiniMed 640G insulin pump (MMT-1512) is approximately 91.9
grams.
The mass of the MiniMed 640G insulin pump (MMT-1712) is approximately 95.7
grams.
■
product specifications and safety information
Product specifications and safety information 259
MP6025958-022 / A
RELEASED