User's Manual

Basal settings
Item Default setting Limits Increments
Max Basal Rate 2 U/h 0–35 U/h 0.025 U for
0.025–0.975 U/h
0.05 U for 1.00–9.95 U/h
0.1 U for rates of 10.0 U/h
or more
Basal Rate 0.000 U/h 0.000 U/h to Max
Basal Rate setting
0.025 U for
0.025–0.975 U/h
0.05 U for 1.00–9.95 U/h
0.1 U for rates of 10.0 U/h
or more
Temp Basal Type Percent Percent, Rate N/A
Temp Basal
Percent
100% 0-200% 5%
Temp Basal Rate Current basal
rate
0.0 U/hr to Max
Basal Rate
0.025 U for
0.025–0.975 U/h
0.05 U for 1.00–9.95 U/h
0.1 U for rates of 10.0 U/h
or more
Insulin sensitivity factor
Maximum settings 8
Default None. Insulin sensitivity is set during Startup of the Bolus
Wizard.
Range 5 to 400 mg/dL/unit
Low Reservoir reminder
The values are based on displayed amount, not actual amount.