User's Manual
• Operating humidity range of the case: 20% to 90%.
This requirement exceeds IEC 60601-1, subclause 7.9.3.1 (30% to 75%).
Filling the infusion set and cannula
• The cannula can be filled from 0.025 units to 5.1 units, in increments of
0.025 units.
• The standard fill rate is 1.5 units per minute.
The quick fill rate is 15 units per minute.
• When filling the tubing, a warning occurs at 30 units, and thereafter at every
10 units.
• Insulin used to fill the infusion set is recorded in the Daily History.
Infusion pressure
The maximum infusion pressure and occlusion pressure is 86.12 kPa (12.49 psi).
Insulin delivery default settings
Bolus settings
Item Default setting Limits Increments
Bolus Wizard
feature:
Off - -
Easy bolus: Off - -
Easy bolus step
size:
0.1 U 0.1 U to 2 U -
Bolus increment: 0.10 U 0.025 U
0.05 U
0.10 U
-
Dual/Square
bolus:
Off - -
Max bolus: 10 U 0 to 75 U (per
single bolus)
-
Bolus BG Check
Reminder:
Off 0:00 to 5:00 0:30
256 Chapter 15
MP6025958-022 / A
RELEASED