User's Manual

Title and text Explanation Next steps
Cannot connect device
This device is
incompatible with your
pump. See User Guide.
You may be trying
to connect a
device that is not
compatible with
your pump.
You are trying to
connect a
transmitter to your
pump but another
transmitter is
already wirelessly
connected to your
pump.
Select OK to clear the alert.
Check the list of devices that are
compatible with your pump in
Optional devices, on page 14.
If you are replacing your
transmitter, make sure that you
first delete the old transmitter
from your pump before you try
to connect your new transmitter.
Only one transmitter can be
connected to your pump. For
details, see Deleting the
transmitter from your pump, on
page 181.
Check settings
Startup Wizard settings
complete. Check and set
up your other settings.
Some settings have
been cleared or
reverted to factory
default values.
Select OK to clear the alert.
Review any settings that you
have not already set in Startup
Wizard and re-enter the values, if
necessary.
Critical pump error
Delivery stopped. Pump
not working properly.
Stop using pump.
Remove infusion set from
body. Consider other
insulin treatment. See
User Guide.
Your pump has
encountered a
critical error.
The pump is not able to deliver
insulin. Remove your infusion set
and stop using your pump.
Consider another form of insulin
delivery.
Check your BG, and treat as
necessary.
Write down the error code that
appears on the alarm screen.
Call your local representative for
assistance with your pump.