User's Manual

Title and text Explanation Next steps
Auto Suspend
Insulin delivery
suspended. No buttons
pressed within time set in
Auto Suspend.
You have not
pressed any buttons
during the time
specified in the
Auto Suspend
settings.
To clear the alarm and resume
basal insulin delivery, select
Resume Basal.
Check your BG and treat as
necessary.
Battery failed
Insert a new AA battery.
The pump battery
does not have
enough power.
Select OK to clear the alarm.
Remove the old battery and
insert a new AA battery.
For details, see About batteries, on
page 21.
Battery not compatible.
See User Guide.
The battery that you
inserted into the
pump is not
compatible.
To clear the alarm, remove the
incompatible battery.
Insert a new AA battery. For
compatible battery types, see
About batteries, on page 21.
Bolus not delivered.
Bolus entry timed out
before delivery. If bolus
intended, enter values
again.
Bolus values
entered, but bolus
was not delivered
within 30 seconds.
Select OK to clear the alert.
If bolus delivery was intended,
check your BG, re-enter bolus
values and deliver bolus.
Bolus stopped
Cannot resume bolus or
cannula fill. XX.XXX of
YY.YYY U delivered.
ZZ.ZZZ U not delivered. If
needed, enter values
again.
The battery power
was exhausted
while a bolus or Fill
Cannula was in
progress.
Note the amount of insulin not
delivered.
Replace the AA battery.
Select OK to clear the alarm.
Deliver the remaining bolus
amount if needed.
alarms, alerts, and messages
Alarms, alerts, and messages 203
MP6025958-022 / A