Note: If you are using Alert Silence and currently have all sensor alerts
silenced, your pump does not display the troubleshooting screens. Any
glucose alerts you received are displayed in the Alarm History screen.
1. Make sure your setup meets the following requirements:
• Your transmitter is fully charged.
If both lights on the charger are off, your transmitter is fully charged. For
details, see your transmitter user guide.
• You have only one transmitter connected to your pump.
Delete the current transmitter that is connected to your pump before
continuing. For details, see Deleting the transmitter from your pump, on
page 181.
• Your transmitter is placed next to your pump.
• Your transmitter is reconnected to the pump, if your pump has been
recently reset. For details, see Wirelessly connecting your pump and
transmitter using Auto Connect, on page 175.
• The Airplane Mode is turned off on your pump.
• You have applied the tape correctly, as instructed in the sensor user
guide.
2. Disconnect the transmitter from the sensor for at least 10 seconds.
3. While the light on the transmitter is blinking, reconnect the transmitter with
the sensor to restart communication and select OK on the pump to
acknowledge the alert.
4. Depending on whether the light blinked when you connected the transmitter
to the sensor, select Yes or No on your pump and do one of the following:
■
troubleshooting
Troubleshooting 235