What is a Check Settings alarm?
This alarm occurs when a condition causes your pump to reset to factory settings.
You see this alarm after your pump guides you through re-entering the Startup
Wizard settings.
The Check Settings alarm is letting you know that other settings may have been
cleared or reverted to factory default values. Review any settings that you have not
already set in Startup Wizard and re-enter the values, if necessary.
My pump is asking me to rewind
WARNING: Make sure the infusion set is disconnected from your
body before you rewind your pump or fill the infusion set tubing.
Never insert the reservoir into the pump while the tubing is
connected to your body. Doing so could result in an accidental
infusion of insulin, which may cause low BG.
You always rewind your pump when changing the reservoir. Rewinding returns the
piston in the reservoir compartment to its starting position. It is normal for your
pump to ask you to rewind anytime you must remove and replace the reservoir,
such as when resolving an Insulin Flow Blocked alarm or addressing a problem
loading the reservoir.
I dropped my pump
Caution: If your pump has been dropped or you suspect your pump
is damaged, carefully inspect your pump to ensure there are no cracks
before exposing your pump to water.
Do the following:
1. Check that all connections are still tightly in place.
2. Check the display, button area, and pump case for cracks or damage.
3. Check the infusion set, including the tubing connector and tubing for cracks
or damage.
4. Review the status screen, basal rates and other pump settings.
5. Perform the Self Test procedure by going to: