User's Manual

Messages
A message informs you about the status of your
pump or if you need to make a decision.
When a message occurs:
Display: The pump displays a notification with a
blue icon and instructions.
Notification light: Does not illuminate or blink.
Audio: Depending on the message, the pump emits a message tone, an alert
tone, or no tone. Depending on your Audio Options settings, you may hear a tone,
feel a one-pulse-only vibration, or hear a tone and feel a vibration.
You clear the message by pressing
and make a selection.
Pump alarms, alerts, and messages
The following table lists the most common or serious alarms, alerts, and messages
related to your pump. The table also explains the meaning, consequences, and the
reasons why these notifications appear, and provides steps for problem resolution.
Title and text Explanation Next steps
Active Insulin cleared
Any Active Insulin amount
has been cleared.
An alarm has
cleared your active
insulin amount.
Your active insulin
amount is now at
0 units.
Select OK to clear the alarm.
The active insulin tracked prior to
pump restart is not included in
new Bolus Wizard calculations.
Consult your healthcare
professional for how long you
need to wait after active insulin is
cleared before you can rely on
the active insulin calculation of
your Bolus Wizard
You can check Daily History for
the time and amount of your last
bolus. For more information, see
Daily History, on page 121.
202 Chapter 12
MP6025958-022 / A
RELEASED