User's Manual

Title and text Explanation Next steps
Rise Alert
Sensor glucose rising
rapidly.
Your SG value has
been rising as fast
or faster than your
preset Rise Alert
Limit.
Select OK to clear the alert.
Monitor trend and glucose level.
Follow instructions from your
healthcare professional.
Sensor alert occurred
Check Alarm History for
silenced alerts.
Sensor alert
occurred when Alert
Silence is on.
Select OK to clear the alert.
Check the Alarm History screen
to see which alerts were silenced.
For more information about
accessing the Alarm History
screen, see Alarm History, on
page 122.
Select the alert to open the
Alarm Detail screen.
Take action based on the
selected alert.
Sensor connected
If new sensor, select Start
New. If not, select
Reconnect.
The transmitter has
detected that you
have connected a
sensor. The pump
needs to know if
this is a new sensor
or if you have
reconnected your
old sensor.
If you have connected a new
sensor, select Start New Sensor.
If you have reconnected a sensor
you have been using, select
Reconnect Sensor.
In either case, a "warm-up"
messages appears on your Home
screen, and you are prompted to
calibrate your sensor. Your pump
starts receiving your SG values
again after the two-hour
initialization is complete.
Sensor expired
Insert new sensor.
The sensor has
been used for
6 days (144 hours).
It has reached the
end of its useful life.
Change your sensor. For details,
see your sensor user guide.
alarms, alerts, and messages
Alarms, alerts, and messages 223
MP6025958-022 / A