Title and text Explanation Next steps
No calibration occurred
Confirm sensor signal.
Calibrate by XX:XX.
The transmitter was
unable to receive
the calibration BG
meter readings from
the pump.
• Select OK to clear the alert.
• Check the status bar on your
pump to ensure that your pump
has a signal from your sensor. If
there is no sensor signal, see My
pump cannot find the sensor
signal, on page 234.
• Calibrate again by the time
shown on the pump screen to
ensure you continue SG
monitoring.
No calibration occurred
Confirm sensor signal.
Check BG again to
calibrate sensor.
The transmitter was
unable to receive
the required
calibration BG from
the pump.
Calibration is
required by the
system for SG
values to resume.
"Calibration
required" appears
on your sensor
graph.
• Select OK to clear the alert.
• Take another BG meter reading
and calibrate again.
Possible signal
interference
Move away from
electronic devices. May
take 15 minutes to find
signal.
There may be
interference from
another electronic
device that is
affecting the
communication
between your
pump and
transmitter.
• Move away from other electronic
devices. It can take up to
15 minutes for your pump to
start communicating with your
transmitter.
• Select OK to clear the alert.
222 Chapter 12
MP6025958-022 / A
RELEASED