User's Manual
Title and text Explanation Next steps
Check connection
Ensure transmitter and
sensor connection is
secure, then select OK.
The pump fails to
detect the
transmitter and is
unable to receive
sensor signal.
• Select OK to clear the alert.
• If your sensor is fully inserted,
select Yes. If your sensor is not
fully inserted, select No.
• If your sensor was not fully
inserted, insert a new sensor.
• If you still cannot connect your
sensor, see My pump cannot find
the sensor signal, on page 234.
Lost sensor signal
Move Pump closer to
transmitter. May take
15 minutes to find signal.
Transmitter signal
has not been
received for
30 minutes during
or after initialization.
• Move your pump closer to your
transmitter. It can take up to
15 minutes for your pump to
start communicating with your
transmitter.
• Select OK to clear the alert.
Low battery transmitter
Recharge transmitter
within 24 hours.
The battery in the
transmitter needs to
be recharged within
24 hours.
• Select OK to clear the alert.
• Recharge your transmitter as
soon as possible.
Medical device
CALL FOR EMERGENCY
ASSISTANCE. I have
diabetes.
Your pump is
suspended due to
low SG, and you
have not responded
to the alarm within
10 minutes.
• Select Dismiss.
• Immediately call for emergency
assistance.
■
alarms, alerts, and messages
Alarms, alerts, and messages 221