User's Manual
Title and text Explanation Next steps
Cannot find sensor
signal
Disconnect and reconnect
transmitter, then select
OK. Notice if transmitter
light blinks.
The pump has not
received a signal
from the
transmitter.
• Disconnect and reconnect your
transmitter and sensor.
• See if the light on your
transmitter blinks when
connected to the sensor. You
may need this information for
troubleshooting later.
• Select OK. Your pump searches
for your sensor. If your pump
receives a signal from your
sensor, you do not need to do
anything else. If your pump does
not receive a signal from the
sensor, another message appears
to let you know.
Change sensor
Insert new sensor and
Start New Sensor.
You selected No in
the Check sensor
insertion message,
indicating that your
sensor is not fully
inserted.
• Select OK to clear the alert.
• Change your sensor. For details,
see your sensor user guide.
• After you change your sensor,
refer to Starting the sensor, on
page 182.
Change sensor
Second calibration not
accepted. Insert new
sensor.
This alert occurs
when you receive
two Calibration not
accepted errors in a
row.
• Select OK to clear the alert.
• Change your sensor. For details,
see your sensor user guide.
Change sensor
Sensor not working
properly. Insert new
sensor.
Sensor signal is no
longer reliable.
• Select OK to clear the alert.
• Change your sensor. For details,
see your sensor user guide.
220 Chapter 12
MP6025958-022 / A
RELEASED