Title and text Explanation Next steps
Calibrate now
Check BG and calibrate
sensor.
A BG meter reading
is needed
immediately to
calibrate your
sensor so that you
can continue
receiving sensor
glucose readings.
• Take a BG meter reading and
enter for calibration. After you
receive a Calibrate now alert, it
takes up to 15 minutes after you
calibrate for your pump to
receive SG values. For details, see
Calibrating your sensor, on
page 183.
• If you are unable to calibrate
now, you can use the Snooze
feature. Set the desired time, and
select Snooze. If you do not
calibrate before the Snooze time
is up, the Calibrate Now alert
occurs again.
Calibration not accepted
Recheck BG and calibrate
sensor.
Your system was
unable to use the
BG meter readings
you entered to
calibrate your
sensor.
• Wash and dry hands thoroughly.
See Guidelines for calibrating, on
page 186.
• Enter a new BG meter reading for
calibration as instructed in
Calibrating your sensor, on
page 183. If you receive a
Calibration not accepted alert on
your second calibration, a
Change sensor alert occurs.
• Call your local representative if
you have questions.
■
alarms, alerts, and messages
Alarms, alerts, and messages 219