User's Manual

Title and text Explanation Next steps
Wrong key pressed
Try again
Wrong button
selected while
unlocking the
pump.
Select OK to clear the alert.
Try to unlock the pump. Select
the highlighted key when
prompted.
For more information, see
Unlocking your pump, on page 26.
CGM (sensor) alarms, alerts, and messages
The following table lists the most common or serious alarms, alerts, and messages
related to your sensor glucose readings, as well as the status of your transmitter
and sensor. The table also explains the meaning, consequences, and the reasons
why these notifications appear, and provides steps for problem resolution.
Title and text Explanation Next steps
Alert before high
Sensor glucose
approaching High Limit.
Check BG.
Your SG value is
approaching your
specified high limit.
Select OK to clear the alert.
Check your BG.
Follow instructions from your
healthcare professional and
continue to monitor your BG.
Alert before low
Sensor glucose
approaching Low Limit.
Check BG.
Your SG value is
approaching your
specified low limit.
Select OK to clear the alert.
Check your BG.
Follow instructions from your
healthcare professional and
continue to monitor your BG.
Alert on high XXX mg/dL
High sensor glucose.
Check BG.
Your SG value is at
or above your
specified high limit.
Select OK to clear the alert.
Check your BG.
Follow instructions from your
healthcare professional and
continue to monitor your BG.
216 Chapter 12
MP6025958-022 / A