User's Manual
Title and text Explanation Next steps
Resume Dual bolus?
XX of YY U delivered.
Resume delivery of ZZ U
now, and AA U Square for
XX:XX hr?
The Now portion of
a Dual Wave bolus
delivery has been
interrupted because
the pump battery
was removed. If it is
within 10 minutes
since this
interruption, you
can resume this
bolus.
• Check the message to see how
much of the Dual Wave bolus
was actually delivered.
• To cancel remaining amount of
bolus, select Cancel.
• To resume remaining amount of
bolus, select Resume.
Resume Square bolus?
XX of YY U delivered for
XX:XX hr. Resume delivery
of ZZ U for XX:XX hr?
The Square Wave
bolus delivery was
interrupted. If it is
within 10 minutes
since this
interruption, you
can resume this
bolus.
• Check the message to see how
much of the Square Wave bolus
was actually delivered.
• To cancel remaining amount of
bolus, select Cancel.
• To resume remaining amount of
bolus, select Resume.
Stuck button
Button pressed for more
than 3 minutes
The pump has
detected that a
button has been
pressed for an
unusually long time.
• Select OK to clear the alarm.
• If this alarm occurs again, contact
your local representative for
assistance with your pump.
If you are unable to clear the
alarm:
• Consider another form of insulin,
because your pump is not
delivering insulin.
• Check your BG and treat as
necessary.
• Contact your local representative
for assistance with your pump.
■
alarms, alerts, and messages
Alarms, alerts, and messages 215
MP6025958-022 / A