Title and text Explanation Next steps
Resume Dual bolus?
XX of YY U delivered.
Resume delivery of ZZ U
now, and AA U Square for
XX:XX hr?
The Now portion of
a Dual Wave bolus
delivery has been
interrupted because
the pump battery
was removed. If it is
within 10 minutes
since this
interruption, you
can resume this
bolus.
Check the message to see how
much of the Dual Wave bolus
was actually delivered.
To cancel remaining amount of
bolus, select Cancel.
To resume remaining amount of
bolus, select Resume.
Resume Square bolus?
XX of YY U delivered for
XX:XX hr. Resume delivery
of ZZ U for XX:XX hr?
The Square Wave
bolus delivery was
interrupted. If it is
within 10 minutes
since this
interruption, you
can resume this
bolus.
Check the message to see how
much of the Square Wave bolus
was actually delivered.
To cancel remaining amount of
bolus, select Cancel.
To resume remaining amount of
bolus, select Resume.
Stuck button
Button pressed for more
than 3 minutes
The pump has
detected that a
button has been
pressed for an
unusually long time.
Select OK to clear the alarm.
If this alarm occurs again, contact
your local representative for
assistance with your pump.
If you are unable to clear the
alarm:
Consider another form of insulin,
because your pump is not
delivering insulin.
Check your BG and treat as
necessary.
Contact your local representative
for assistance with your pump.